This signal analyzing equipment enables us to conduct a simultaneous measurement for frequency response generated from loud speaker, SPL, impedance, Fo, Ze, THD, Phase, polarity, and Rub & Buzz. High-frequency signal analysis offers even more accurate testing and identification of Rub&Buzz than existing hearing tests. The algorithm of signal analysis applied to this device allows to detect even minute Rub & Buzz due to its innovative and new method. The 4-stage defect setting enables even more in-depth assessment.
Special Feature :
1. Simultaneous measurement and assessment of Frequency Response (SPL), Impedance, Fo, Ze, THD, Phase, Polarity, and Rub&Buzz.
2. High performance SNR>100dB.
3. Sampling rate up to 48kHz
4. Auto calibration : Microphone. Output voltage.
5. 4-stage pass/fail grade setting and management
6. Self-test supported for the hardware system.
7. Computer-controlled operation Fast, hot plug USB interface.
8. Records measurement data in a file, calculate and conduct statistical processing (SPC) regarding the Avg, Min, Max, Cp, and Cpk of SPL, Fo, Ze, THD (4-point), Rub & Buzz (4-point).
9. Can perform remote control through the Internet.
10. Easy document printing in MS Word or Excel file (text format only)
11. MIC Pre-Amp, Conditioning Amp, Power Amp System control. Remote control.
12. As the dual input and output terminals allow a synchronized measurement, a stereo set can be analyzed with one time measurement.
* Option-01A : Once connected to the automation system, outputs separate signals divided into 4 classes of pass or fail for all measurement items.
SG-5622AA Technical Specifications
Generator & Analyzer
Level flatness of the combined generator & analyzer
100 Hz ~ 10kHz
20 Hz ~ 20 kHz
Sweep Generator
Number of channels
Sampling rate (fs) / Level resolution
Frequency range
Internal frequency accuracy
Selectable input impedance per channel
Input impedance accuracy
Input impedance coupling
Phantom power voltage
Phantom voltage accuracy
Phantom power inward resistance
15V
Measurements
Option & Accessories
Phanton power voltage
Polarization Voltage
SP-72
SG-5001B
General
Standard software
Interface Bus
Main power requirements
Fuse
Operating temperature
Dimensions
Weight
<± 0.1 dB
<± 0.2 dB
2 channels with individual levels
48 kHz / 24 bit DAC
20 Hz ~ 20 kHz
0.1 dB
94 dB
<± 0.1 dB
<± 0.1 dB
<± 0.2 dB
Log : 1/1,1/3,1/6,1/12,1/24,1/48 Octave steps
Lin : 1 to 2000 steps
<± 50 ppm
10mVRMS to 10.0VRMS , 5W [Max]
0.1V (8Ω) at 1 kHz -60 dB
1V to 5V (8Ω) at 1 kHz -65 Db
2 independent channels with individual input range
48 kHz / 24 bit
20 Hz ~ 20 kHz
-34 to + 60 dBVp
<± 0.1 dB
<± 0.1 dB
<± 0.2 dB
<-85 dB
<-120 dB
± 1 deg @ 200 Hz ~ 4 kHz
± 50 ppm
1 MΩ
1% of resistor value
AC Coupled
15 VDC
± 0.5 VDC
2.2 KΩ
[1] SPL (Sound Pressure Level) Frequency Response
[2] Total Harmonic Distortion / 2nd & 3rd Harmonisc Distortion
[3] Extraneous Noise (Rub&Buzz) Frequency Response
[4] Impedance Frequency Response
[5] Polarity (Phase Frequency Response)
[6] Impedance
[7] Fo, Ze (Resonance Frequency)
[8] L/R Comparison measurement
Connector type BNC 24 VDC, 48 VDC
Connector type 7 pin Lemo 200 VDC
Lemo ICP Adaptor
Remote Control
15P D-Type connector I/O signal - Start, Pass, Fail, Fail [Class1, Class2, Class3, Class4]
SPL Analyzer ON. Rub & Buzz Analyzer ON .
Remote control I/O ※ Terminal Box [Option 01A]
Audio Analyzer Production software
(PC based Windows® XP, Vista, win7 )
Pentium®4, 1.5GHz processor minimum or equivalent.
(Celeron processors are not recommended)
Personal computer connected to the instrument
through the Audio Analyzer Interface Bus. (USB2.0)
AC 220-230V~ 50/60Hz 95VA
5 x 20 T LBC 1A/250V Standard : EN10480
+ 5 to +45℃ ( +40 to +110 F ), R.H.<90% non condensing
485(W) * 95(H) * 256(D) mm (without Dimensions of Power Amp)